Ultrasonic microdialysis coupled with capillary electrophoresis electrochemiluminescence study the interaction between trimetazidine dihydrochloride and human serum albumin.

Article Details

Citation

Sun S, Long C, Tao C, Meng S, Deng B

Ultrasonic microdialysis coupled with capillary electrophoresis electrochemiluminescence study the interaction between trimetazidine dihydrochloride and human serum albumin.

Anal Chim Acta. 2014 Dec 3;851:37-42. doi: 10.1016/j.aca.2014.08.012. Epub 2014 Aug 13.

PubMed ID
25440662 [ View in PubMed
]
Abstract

The paper describes a homemade ultrasonic microdialysis device coupled with capillary electrophoresis electrochemiluminescence (CE-ECL) for studying the interaction between human serum albumin (HSA) and trimetazidine dihydrochloride (TMZ). The time required for equilibrium by ultrasonic microdialysis was 45min, which was far less than that by traditional dialysis (240min). It took 80min to achieve the required combination equilibrium by normal incubation and only 20min by ultrasonic. Compared with traditional dialysis, the use of ultrasonic microdialysis simplified experimental procedures, shortened experimental time and saved consumption of sample. A simple, sensitive and selective determination of TMZ was developed using CE-ECL and the parameters that affected ECL intensity were optimized. Under the optimized conditions, the linear range of TMZ was from 0.075 to 80mumol/L (r(2)=0.9974). The detection limit was 26nmol/L with RSD of 2.8%. The number of binding sites and binding constant were 1.54 and 15.17L/mol, respectively.

DrugBank Data that Cites this Article

Drugs
Drug Carriers
DrugCarrierKindOrganismPharmacological ActionActions
TrimetazidineSerum albuminProteinHumans
Unknown
Binder